PUBLISHED

CVE-2025-0626

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor



Description

Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.

Reserved 2025-01-21 | Published 2025-01-30 | Updated 2025-01-31 | Assigner icscert


HIGH: 7.7CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N

HIGH: 7.5CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H

Problem types

CWE-912 Hidden Functionality

Product status

Default status
unaffected

All versions
affected

Credits

An anonymous researcher reported these vulnerabilities to CISA. finder