CVE-2025-0626

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor



Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Reserved 2025-01-21 | Published 2025-01-30 | Updated 2025-03-01 | Assigner icscert


HIGH: 7.7CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N

HIGH: 7.5CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H

Problem types

CWE-912 Hidden Functionality

Product status

Default status
unaffected

All versions
affected

Credits

An anonymous researcher reported these vulnerabilities to CISA. finder

References

www.cisa.gov/...vents/ics-medical-advisories/icsma-25-030-01

www.fda.gov/...s-contec-and-epsimed-fda-safety-communication

cve.org (CVE-2025-0626)

nvd.nist.gov (CVE-2025-0626)